CSO Anterior Segment OCT MS-39

MS-39 is the most advance device for the analysis of the anterior segment of the eye. The Anterior Segment OCT MS-39 combines Placido disk corneal topography and a high resolution OCT-based anterior segment tomography. Available in the U.S.A.

 DOWNLOAD CATALOG     PRODUCT MANUAL

Description

MS-39 is known to produce the highest resolution and clarity of the cross-sectional images, with a 16 mm diameter, along with the many details of the cornea structure and layers. MS-39 can measure and detect on the Placido image so that height, slope, and curvature data can be calculated using the arc-step method with conic curves. Profiles of the anterior cornea, posterior cornea, anterior lens, and iris are derived from the SD-OCT scans.

Features

  • Can be used in corneal surgery for refractive surgery planning
  • Uses the Phoenix software platform
  • USB connection for fast and easy transfer
  • High precision joystick
  • Provides information on pachymetry, elevation, curvature and dioptric power
  • Optional Electrical Tables
  • Pinpoints the clarity of the cross-sectional images with a 16 mm diameter
  • Offers a complete overview of the corneal aberrations
  • Advanced measurement of the epithelial layer
  • Provides an accurate calculation of the spherical and toric power
  • Enables the measurement of AOD, TISA and corneal pachymetry
  • Optional Tabletop and Monitors

Software

  • Epithelial map
  • Crystalline biometry
  • Advanced analysis of the tear film
  • Pupillograph
  • Keratoconus screening
  • IOL calculation module
  • Intrastromal ring
  • Glaucoma screening
  • Corneal aberrometry

Topographic Maps

  • Tangential curvature (anterior and posterior)
  • Sagittal curvature (anterior and posterior)
  • Elevation (anterior and posterior
  • Refractive power (equivalent, anterior and posterior)
  • Corneal thickness
  • Epithelial thickness
  • Anterior chamber depth

Technical Data

DATA TRANSFER USB 3.0
POWER SUPPLY External power source 24 VCC
POWER NET CABLE IEC C14 plug
CHINREST MOVEMENT 70mm ± 1mm
CHIN CUP HEIGHT 23cm
BASE MOVEMENT 105 x 110 x 30 mm
WORKING DISTANCE 74mm
DIMENSION 505 x 315 x 251mm
WEIGHT 10.4Kg

Light Sources

PLACIDO’S LED LIGHTING Led @635nm
OCT SOURCE Led @845nm
PUPILLOGRAPHIC ILLUMINATION Led @950nm

Topography

PLACIDO DISK RINGS 22
MEASURED POINTS 31232 (anterior surface) 25600 (posterior surface)
TOPOGRAPHIC COVERING from 1D to 100D
MEASUREMENT ACCURACY Class A according to UNI EN ISO 19980-2012

Topography

IMAGE FIELD 16mm x 8mm
IMAGE(S) RESOLUTION Keratoscopy (640x480) + 25 radial scans on a 16mm transversal field
AXIAL RESOLUTION 3.6 µm (in tissue)
TRANSVERSAL RESOLUTION 35 µm (in air)
OPERATING SYSTEM Windows 10 (64 bit)

Video Demo

How to find us

Viewlight is a leading provider of medical practice equipment for ophthalmologists, optometrists and opticians.

Ask a question

    Additional information
    Brand

    CSO

    Country of Origin

    Made in Italy

    FDA

    Approved

    Optional Accessories

    Double Item Work Station

    ,

    Single Item Work Station

    Reviews (0)

    Reviews

    There are no reviews yet.

    Only logged in customers who have purchased this product may leave a review.

    About CSO
    Shipping and Receiving

    SHIPPING & DELIVERY

    Orders are shipped Monday through Friday, please contact your sell representative or customer service for exact information.

    If depend on many factors , due instruments needs to be inspected and tested before shipping,  please contact your sell representative or customer service for exact information.

    Contact customer service immediately. Sometimes Cancellations are no longer possible as your order will already have been sent to our fulfillment center. You will have to wait until the item is delivered and then return it for a full refund.

    Unfortunately, we cannot modify an order once it has been placed. We offer return policy if you'd like to return any unwanted products once they've been delivered, please contact customer service immediately.

    These policies only apply within the United States. If you are not in the United States, we don’t accept any returned or cancellation.

    All orders are processed within 3-7 business days. Orders are not shipped or delivered on weekends or holidays. If we are experiencing a high volume of orders, shipments may be delayed by a few days. Please allow additional days in transit for delivery. If there will be a significant delay in shipment of your order, we will contact you via email or telephone.

    Viewlight USA Full Shipping Policies: SEE ALL

    Shipping Services
    About Brand

    Viewlight USA is engaged in the development, sale, and maintenance for most of the ophthalmic and optical instruments. Viewlight USA is established and located in Miami, FL. We offer top of the line equipment with the latest medical technologies and a commitment to provide the finest medical equipment in the market.

    We are proud to partner up with leading optics brands: CSO | Inami & Co., Ltd | Haag-Streit.